Cable IQ Internal Errors
 
22 - Oscillator fail. Most likely a hardware failure and will need to go in for repair.
 
23 - Invalid data address Return for repair if the error persists
 
24 - Stack error. Notify [email protected]
 
25 - Math error. Notify [email protected]
 
31 - Synthesizer not locked. Most likely a hardware failure and will need to go in for repair.
 
32 - Calibration record error. Return for repair.
 
33 - Invalid A/D channel used. Notify [email protected]
 
34 - Flash error. Return for repair if error persists
 
35 - DO LOOP nesting error. Notify [email protected].
 
36 - Serial Flash CRC error, data is missing or corrupted. Return for repair if error persists after autotests have been erased. [email protected]
 
41 - Invalid language requested. Notify [email protected]
 
42 - HI GAIN error. Return for repair if error persists.
 
43 - Data error. Notify [email protected]
 
44 - Enque/Deque error. Notify [email protected]
 
45 - Clock error. Notify [email protected] and return for repair if error persists.